Standardising and validating aromatic profile analysis by an electronic nose

Estandarización y validación del método de análisis del perfil aromático por nariz electrónica

DOI:

https://doi.org/10.15446/ing.investig.v31n2.23466

Keywords:

Repeatability, intermediate repeatability, uncertainty, aromatic profile, food. (en)
repetibilidad, repetibilidad intermedia, incertidumbre, perfil aromático, alimentos. (es)

Authors

  • Carlos Mario Zuluaga Universidad Nacional de Colombia
  • Amanda Consuelo Díaz Universidad Nacional de Colombia
  • Martha Cecilia Quicazán Universidad Nacional de Colombia

The electronic nose is one of the most innovative techniques which had been used in the analysis of food and other aroma matrices. This device imitates the human olfactory system through sensors which are partially specific for different groups of volatile chemicals and differentiating and classifying food (by involving multivariate statistics) and it is also used for establishing quality and control parameters. Although the technique has been increasingly applied in research and industry, few studies have been published in the scientific literature referring to validating its use as an analytical methodology. This work' s main objective was to standardise and validate the aromatic profile analysis method for differentiating apiculture products: honey, pollen and propolis. Standardisation involved modifying sample preparation and system variables while validation involved analysis of repeatability and intermediate repeatability of the method and establishing its uncertainty. The results showed that it was possible to set suitable operational parameters for each apicultural product while validation data showed that this technique did not reveal significant differences as reliable values were obtained when using this device as an analysis and quality control tool.

Una de las técnicas más novedosas en el análisis del aroma de alimentos y otras matrices es la nariz electrónica. Este dispositivo funciona como una imitación del sistema olfatorio humano por medio de sensores con sensibilidades solapadas para diferentes grupos de compuestos químicos volátiles que, mediante el uso de estadística multivariada, permite la diferenciación y clasificación de alimentos así como el establecimiento de parámetros de calidad y control de los procesos. Aunque la técnica cada día es más utilizada en la investigación y en la industria, son pocos los trabajos publicados en la literatura científica que hacen referencia a la validación de su uso como metodología analítica. El objetivo de este trabajo fue el de estandarizar y validar el método de análisis del perfil aromático por nariz electrónica para la diferenciación de productos apícolas: miel, polen y propóleo. La estandarización se logró modificando las condiciones de preparación de la muestra y las variables del sistema, mientras que la validación se realizó mediante el análisis de la repetibilidad, la repetibilidad intermedia del método y el cálculo de la incertidumbre. Los resultados obtenidos muestran que es posible determinar los parámetros adecuados de operación del instrumento para cada uno de los productos apícolas, mientras que los indicadores de validación mostraron que la técnica no presenta diferencias significativas en su respuesta, obteniéndose valores confiables para el uso de este equipo como una herramienta de análisis y control de calidad.

Estandarización y validación del método de análisis del perfil aromático por nariz electrónica

Standardising and validating aromatic profile analysis by an electronic nose

Carlos Mario Zuluaga1, Amanda Consuelo Díaz2, Martha Cecilia Quicazán3

1 Ingeniero Químico, Magister en Ciencia y Tecnología de Alimentos, Universidad Nacional de Colombia. Estudiante de doctorado de Ingeniería Química, Universidad Nacional de Colombia, Colombia. cmzuluagad@unal.edu.co

2 Ingeniera de Alimentos, Universidad de la Salle. Doctora en Calidad, Seguridad y Tecnología de Alimentos, Universidad de Zaragoza, España. Profesora Asistente, Universidad Nacional de Colombia. amcdiazmo@unal.edu.co

3 Ingeniera Química, Universidad Nacional de Colombia. Magister en Ciencia y Tecnología de Alimentos, Universidad de La Habana, Cuba. Profesora Asociada, Universidad Nacional de Colombia. mcquicazand@unal.edu.co


RESUMEN

Una de las técnicas más novedosas en el análisis del aroma de alimentos y otras matrices es la nariz electrónica. Este dispositivo funciona como una imitación del sistema olfatorio humano por medio de sensores con sensibilidades solapadas para diferentes grupos de compuestos químicos volátiles que, mediante el uso de estadística multivariada, permite la diferenciación y clasificación de alimentos así como el establecimiento de parámetros de calidad y control de los procesos. Aunque la técnica cada día es más utilizada en la investigación y en la industria, son pocos los trabajos publicados en la literatura científica que hacen referencia a la validación de su uso como metodología analítica. El objetivo de este trabajo fue el de estandarizar y validar el método de análisis del perfil aromático por nariz electrónica para la diferenciación de productos apícolas: miel, polen y propóleo. La estandarización se logró modificando las condiciones de preparación de la muestra y las variables del sistema, mientras que la validación se realizó mediante el análisis de la repetibilidad, la repetibilidad intermedia del método y el cálculo de la incertidumbre. Los resultados obtenidos muestran que es posible determinar los parámetros adecuados de operación del instrumento para cada uno de los productos apícolas, mientras que los indicadores de validación mostraron que la técnica no presenta diferencias significativas en su respuesta, obteniéndose valores confiables para el uso de este equipo como una herramienta de análisis y control de calidad.

Palabras clave: repetibilidad, repetibilidad intermedia, incertidumbre, perfil aromático, alimentos.

ABSTRACT

The electronic nose is one of the most innovative techniques which had been used in the analysis of food and other aroma matrices. This device imitates the human olfactory system through sensors which are partially specific for different groups of volatile chemicals and differentiating and classifying food (by involving multivariate statistics) and it is also used for establishing quality and control parameters. Although the technique has been increasingly applied in research and industry, few studies have been published in the scientific literature referring to validating its use as an analytical methodology. This work’s main objective was to standardise and validate the aromatic profile analysis method for differentiating apiculture products: honey, pollen and propolis. Standardisation involved modifying sample preparation and system variables while validation involved analysis of repeatability and intermediate repeatability of the method and establishing its uncertainty. The results showed that it was possible to set suitable operational parameters for each apicultural product while validation data showed that this technique did not reveal significant differences as reliable values were obtained when using this device as an analysis and quality control tool.

Keywords: Repeatability, intermediate repeatability, uncertainty, aromatic profile, food.


Recibido: mayo 10 de 2010

Aceptado: julio 17 de 2010

Introducción

Existe una continua necesidad por el desarrollo de métodos analíticos confiables para el análisis de alimentos que puedan ser evaluados con un procedimiento de validación. El Diario Oficial de la Comunidad Europea define validación como: “confirmación mediante examen y puesta a disposición de pruebas efectivas, de que se cumplen los requisitos particulares de un uso específico previsto" (CCE, 2002); en otras palabras, se refiere al proceso de evaluar el rendimiento de un instrumento o metodología cuyo fin último es proveer evidencia objetiva de que el método tendrá una reproducibilidad aceptable y precisa para ser aplicable (Walton, 2001).

Típicamente, cualquier procedimiento analítico debe considerar la siguiente información como un soporte para su evaluación: una descripción del método que incluya detalles de los parámetros operacionales, tales como la preparación de la muestra, el método de extracción, los estándares de referencia y la preparación de reactivos. De igual manera, las características más relevantes cuando se realiza la validación de una metodología analítica son: especificidad, linealidad, precisión, exactitud, rango, límites de detección y cuantificación, y robustez, las cuales pueden ser aplicadas parcial o totalmente dependiendo de la naturaleza de la prueba (APVMA, 2004).

El Instituto de Ciencia y Tecnología de Alimentos (ICTA), de la Universidad Nacional de Colombia, desarrolla el proyecto de investigación Selección de indicadores fisicoquímicos mediante aplicación de nariz electrónica para la catalogación de productos apícolas, financiado por el Ministerio de Agricultura y Desarrollo Rural, cuyo objetivo consiste en mejorar la competitividad de la cadena apícola por medio del establecimiento de las herramientas tecnológicas y científicas que se requieren para caracterizar sus productos, y en un futuro cercano promover un sello de calidad como lo es la denominación de origen para productos de la colmena diferenciados, considerando el gran potencial que tiene esta cadena productiva en el país.

En la primera etapa de desarrollo de este proyecto se estandarizó y validó la metodología de perfil aromático utilizando nariz electrónica para miel, polen y propóleo, cuyos resultados son presentados en este artículo. Una nariz electrónica está formada por cuatro elementos principales: una técnica de extracción de aroma o sistema de flujo de aire que transporta los compuestos volátiles de las muestras al siguiente paso, un arreglo de sensores químicos que transforman el aroma en señales eléctricas, un sistema de instrumentación y control que mide la señal de los diferentes sensores, y el control y automatización del sistema entero (Mannino et al., 2007). El propósito de combinar un arreglo de sensores es proveer una señal global llamada “huella digital” como una respuesta simple que es un aroma característico de la muestra (López-Feria et al., 2008).

El principio básico de funcionamiento de la nariz electrónica consiste en la inyección del headspace (espacio gaseoso ocupado por los compuestos volátiles en el vial) de la muestra colocada en un vial cromatográfico de 20 mL, cuya cantidad no debe exceder en ningún momento más de 1/3 del volumen del vial; la muestra se deja durante un tiempo y temperatura a determinar con el fin de lograr el equilibrio de la concentración de los compuestos volátiles. Una vez inyectado el headspace se obtiene un registro adimensional de relación de conductividades (G/Go) en función del tiempo por cada uno de los sensores de la nariz electrónica, dado por el valor reportado por el sensor en su estado natural (Go) y aquel registrado cuando pasa la sustancia gaseosa a través de él (G) (ver figura 1), la condición esperada es aquella que permita una estabilidad en la respuesta en el menor tiempo.

A pesar de que en la literatura científica ha sido demostrada la utilidad de la nariz electrónica para clasificar y diferenciar de acuerdo con el origen, en el control de procesos y calidad de productos de la industria agroalimentaria (Aleixandre et al., 2008; Ampuero et al., 2004; Benedetti et al., 2004; Cosio et al., 2006; Lammertyn et al., 2004), no se encuentran reportes sobre la estandarización y validación de la metodología de análisis del perfil aromático. La estandarización del método consiste en encontrar aquellos parámetros operacionales y de preparación de la muestra que garanticen una respuesta adecuada y reproducible por parte del equipo. Una vez estos parámetros son hallados, la validación permite evaluar si el método planteado es aplicable; considerando la naturaleza de valoración de esta técnica se estableció que las características más relevantes para la validación son las siguientes: la precisión, la repetibilidad y la repetibilidad intermedia, las cuales garantizan que los datos recolectados son confiables y reproducibles.

La precisión es la capacidad de un instrumento de dar el mismo resultado en diferentes mediciones llevadas a cabo en las mismas condiciones. La repetibilidad (r) está relacionada al rango de valores en los cuales es posible obtener una respuesta del instrumento, al realizarse varias medidas bajo las mismas condiciones de material, equipo, medio ambiente y operador en un período de tiempo corto; su cálculo se realiza mediante la diferencia entre el resultado mayor y el menor del análisis hecho por cada operador y para cada uno de los sensores de la nariz electrónica (ecuación 1) (INS, 2005).

[1]

Entre tanto, la repetibilidad intermedia (R) está relacionada con el rango de valores en los cuales es posible obtener una respuesta del instrumento en el mismo laboratorio pero empleando diferentes operadores, días, equipos o lotes de reactivos; el cálculo es análogo al de repetibilidad, esta vez estableciendo la diferencia entre el resultado mayor y menor de todos los resultados obtenidos (ecuación 2) (INS, 2005).

[2]

Desarrollo experimental

Materiales

Se utilizó una nariz electrónica Airsense Analytics GmbH PEN3 (Schwerin, Alemania); este equipo tiene un arreglo de diez sensores de óxido metálico (MOS), como se muestra en la tabla 1. Las muestras de miel, polen y propóleo fueron entregadas por las asociaciones de apicultores vinculadas al proyecto en mención.

Parámetros de estandarización

Para cada producto apícola se siguió el mismo procedimiento, variando tanto las condiciones propias del producto (peso de muestra) como de operación del equipo (temperatura de acondicionamiento del vial, tiempo de estabilización del headspace y flujo de gas en la cámara de sensores), en tres etapas (figura 2). La primera evaluó el factor de peso de muestra con tres niveles (1, 2 y 3 g), la segunda valoró dos factores: tiempo y temperatura de estabilización del headspace con cuatro niveles cada uno (5, 10, 15 y 20 min) y la tercera consideró el factor de flujo de gas a la cámara de los sensores con tres niveles (60, 100 y 200 mL/ min), y para cada nivel se realizaron tres réplicas; las condiciones más adecuadas se seleccionaron con base en la respuesta más estable y menor variabilidad de los sensores en el tiempo. Una vez obtenido el valor más adecuado en una etapa, se utilizó como parámetro en la siguiente.

Se utilizó como factor de criterio en las etapas de estandarización el valor obtenido del coeficiente de variación (CV), el cual es una medida de dispersión útil para relacionar el promedio con la desviación estándar (s), de acuerdo con la ecuación 3:

[3]

Parámetros de validación

Para la validación de la metodología analítica se evaluó la repetibilidad y la repetibilidad intermedia de cada producto (miel, polen y propóleo), con diversos analistas y en diferentes días; estos parámetros fueron a su vez establecidos para cada uno de los sensores de la nariz electrónica. Se efectuaron cinco réplicas por ensayo; para la evaluación de la repetibilidad intermedia se hizo la prueba t-student con la finalidad de conocer si las diferencias eran estadísticamente significativas entre analista y días distintos, para un valor de significancia α = 0,05, utilizando el paquete estadístico de Microsoft Excel®. Se utilizó la ecuación 4 en el cálculo de la incertidumbre debido a la precisión (U) con los datos de desviación estándar (SD) y el número de veces que se analiza la muestra (N) (Maroto et al., 2003; Reutemann y Huber, 2003):

[4]

Resultados

Estandarización

Los resultados obtenidos en la etapa de estandarización para miel, polen y propóleo se muestran en las tablas 2, 3 y 4 respectivamente, de manera resumida, presentando únicamente los valores máximos y mínimos encontrados a partir de las respuestas de los diez sensores de la nariz electrónica. En miel (tabla 2), la variación en el peso de la muestra (etapa I) permitió establecer que a mayor cantidad de muestra, mayor el coeficiente de variación de respuesta en el último segundo de los sensores, también se observó que la estabilidad de respuesta de éstos a lo largo del tiempo era mayor, con un peso de muestra de 3g, por lo que se utilizó dicha cantidad en las etapas posteriores. La variación de la temperatura y tiempo de acondicionamiento del vial permitió establecer que la mejor respuesta era obtenida con un tiempo de 20 min a 40 °C, y que a pesar de no tener las respuestas de los sensores con la menor variabilidad, sí mostraba una respuesta estable en un tiempo de análisis inferior a las demás; esta etapa también permitió establecer que un tiempo de análisis de 150 s es suficiente para garantizar la estabilidad de respuesta. En la última etapa se encontró que la respuesta es afectada por la cantidad de gas que entra a la cámara; por lo tanto, en la espera de resultados confiables, una menor variablidad es hallada cuando se trabaja con flujos bajos como 60 mL/min.

En polen (tabla 3) la etapa I mostró una señal estable de sensores a lo largo del tiempo con un peso de muestra de 2 g, de manera similar se encontró que la cantidad de muestra afecta considerablemente la repetibilidad en la respuesta. La etapa II permitió deducir que el calentamiento del vial no logra una reducción en el tiempo de análisis de manera considerable, comparada a temperatura ambiente, donde un tiempo de estabilización del headspace de 15 min es suficiente para obtener una respuesta adecuada, con poca variabilidad y en un tiempo corto de inyección, de 220 s. A diferencia de lo encontrado en el análisis de la miel, los flujos altos de gas a la cámara de sensores permiten obtener una respuesta más estable; valores superiores a 200 mL/ min tienen una variabilidad mínima.

En propóleo (tabla 4), la etapa I fue llevada a cabo con pesos de muestra de 1 y 2 g, ya que debido al tamaño de la matriz, cantidades superiores ocupaban más de 1/3 del tamaño del vial, y adicionalmente era poco factible la reducción del tamaño de la muestra, ya que partículas muy pequeñas causan daños severos a los sensores del equipo; se encontró que 2 g dan respuestas estables con menor variabilidad. La etapa II permitió observar que el calentamiento es innecesario para obtener una respuesta de señal estable, pues se obtuvieron valores de relación de conductividades similares a temperatura ambiente, siendo un tiempo de estabilización del vial de 15 min aquel que ofrecía una respuesta de señal menos variable a lo largo del tiempo; en esta misma etapa fue determinado que un tiempo de análisis equivalente a 180 s es suficiente para lograr una estabilidad total para los diez sensores.

Validación

Con los parámetros de análisis determinados en la etapa de estandarización para cada producto, se realizó la validaci ón del método. Los resultados obtenidos en el cálculo del coeficiente de variación y repetibilidad son mostrados en la tabla 5 para miel, polen y propóleo respectivamente; se presentan los rangos de valores máximos y mínimos encontrados a partir de los resultados de los diez sensores de la nariz electrónica.

Entre tanto, en la tabla 6 se presentan el rango de valores obtenidos para la repetibilidad intermedia, incertidumbre y estadístico t para el análisis de promedios, con un valor crítico de ± 2,7764.

En términos generales, se observa que la repetibilidad en el análisis de compuestos volátiles por nariz electrónica tiene una reducida variabilidad. Algunos autores reportan valores de referencia para la evaluación de la precisión del método con base en los coeficientes de variación para métodos cromatográficos y volumétricos menores a 2% (Boqué et al., 2008; Reutemann y Huber, 2003), espectrofotométricos menores a 3% (Perel et al., 1978) y microbiológicos inferiores al 5% (Nguz, 2007). En promedio, los resultados obtenidos para el análisis de perfil aromático con nariz electrónica oscilan entre 7 al 10% para miel, 1 al 4% para polen y 2 al 7% para propóleo; esto permite establecer que las muestras líquidas presentan una mayor variablidad en la respuesta.

Los resultados de la prueba t para el análisis de la precisión entre analistas diferentes y días diferentes muestra para todos los casos que no hay diferencias estadísticas entre ellos, permitiendo establecer que el análisis del perfil aromático de productos apícolas mediante el uso de una nariz electrónica ofrece resultados repetibles en cada uno de los sensores. Los valores de incertidumbre son especialmente bajos para analistas diferentes en polen y propóleo; en miel, la incertidumbre supera el 1% en la mayoría de casos. En días diferentes, la incertidumbre para la respuesta de los sensores en el caso de polen y propóleo vuelve a ser baja, pero en miel los valores superan el 5% en la mayoría de sensores.

Al comparar valores con otros métodos analíticos de alto grado de imprecisión, la metodología es confiable, teniendo en cuenta la especificidad y selectividad de este tipo de análisis en comparación con los métodos cromatográficos capaces de separar compuestos, mientras que esta metodología identifica patrones o “huellas digitales”, las cuales son agrupaciones de sustancias con características comunes, detectados por los sensores mediante un cambio en una relación de conductividades; la nariz electrónica no funciona con respuestas lineales, es decir, sus respuestas están supeditadas a una serie de condiciones establecidas en la estandarización que permiten la respuesta más estable posible en el menor tiempo de análisis.

Conclusiones

La estandarización de la metodología de determinación del perfil aromático en miel, polen y propóleo fue llevada a cabo, encontrándose que las condiciones más adecuadas de operación son:

La validación de las metodologías analíticas permitió establecer que estadísticamente los métodos son precisos y repetibles, mediante el análisis de la medición del perfil aromático en días diferentes y con analistas diferentes; los valores de incertidumbre en la repetibilidad y repetibilidad intermedia para los sensores de la nariz electrónica en miel oscilan entre 1 al 10%, en polen entre 0,5 al 5% y propóleo entre 0,2 al 5%, lo cual permite concluir que la miel es la matriz con la que los sensores ofrecen la respuesta más variable; generalmente las muestras líquidas presentan este comportamiento, mientras que las matrices sólidas tienden a generar una incertidumbre menor. Estos resultados son un paso inicial para futuras aplicaciones de la nariz electrónica en varios campos, especialmente en el de control de calidad; varios laboratorios del mundo actualmente se enfocan en el desarrollo de nuevas tecnologías para la evaluación del comportamiento aromático de sustancias, con el objetivo de encontrar instrumentos que tengan respuestas más precisas, para lo que aún se requiere una mayor investigación.

Agradecimientos

Los autores desean agradecer el apoyo prestado para la realización de este trabajo al Instituto de Ciencia y Tecnología de Alimentos (ICTA), de la Universidad Nacional de Colombia, sede Bogotá; al Ministerio de Agricultura y Desarrollo Rural; a la Asociación de Apicultores de la Región del Sumapaz; a la Asociación de Apicultores de Boyacá; a la Asociación Apícola Comunera; a la Asociación de Apicultores Conservacionistas de la Sierra Nevada de Santa Marta; al Programa Jóvenes Investigadores e Innovadores, del Departamento Administrativo de Ciencia, Tecnología e Innovación (Colciencias); y al Programa de Maestría en Ciencia y Tecnología de Alimentos de la Facultad de Ciencias Agropecuarias de la Universidad Nacional de Colombia, sede Medellín.


Referencias

Aleixandre, M., Lozano, J., Gutiérrez, J., Sayago, I., Fernández, M., Horrillo, M., Portable e-nose to classify different kinds of wine., Sensors and Actuators, B 131, 2008, pp. 71 - 76.

Ampuero, S., Bogdanov, S., Bosset, J., Classification of unifloral honeys with an MS-based electronic nose, using different sampling modes: SHS, SPME and INDEX., European Journal of Food Research and Technology, 218, 2004, pp. 198-207.

APVMA., Guidelines for the validation of analytical methods for active constituent, agricultural and veterinary chemical products., Australian Pesticides & Veterinary Medicines Authority, Kingston, Australia, 2004.

Benedetti, S., Mannino, S., Sabatini, A., Marcazzan, G., Electronic nose and neural network use for the classification of honey., Apidologie 35, 2004, pp. 397-402.

Boqué, R., Maroto, A., Vander, Y., Assessment of accuracy in chromatographic analysis., LCGC Europe 21, 2008, pp. 15- 21.

CCE., Decisión de la Comisión por la que se aplica la Directiva 96/23/CE del Consejo en cuanto al funcionamiento de los métodos analíticos y la interpretación de los resultados., Diario Oficial de la Unión Europea, 2002.

Cosio, M., Ballabio, D., Benedetti, S., Gigliotti, C., Geographical origin and authentication of extra virgin olive oils by an electronic nose in combination with artificial neural networks., Analytica Chimica Acta, 567, 2006, pp. 202-210.

INS., Selección y validación de métodos de ensayo., Instituto Nacional de Salud. Lima, Perú, 2005.

Lammertyn, J., Veraverbeke, E., Irudayaraj, J., Nose technology for the classification of honey based on rapid aromaprofiling., Sensors and Actuators, B 1, 2004, pp. 54 - 62.

López-Feria, S., Cárdenas, S., García-Mesa, J., Valcárcel, M., Simple and rapid instrumental characterization of sensory attributes of virgin olive oil based on the direct coupling headspace-mass spectrometry., Journal of Chromatography A 1188, 2008, pp. 308 - 313.

Mannino, S., Benedetti, S., Buratti, S., Cosio, M., Scampichio, M., Electrochemical sensors for food authentication., Electrochemical Sensor Analysis, Elsevier, 2007, pp. 1091-1098.

Maroto, A., Boqué, R., Riu, J., Rius, X., Cálculo de incertidumbre en medidas químicas: Análisis cromatográfico., Departamento de Química Analítica y Química Orgánica, Instituto de Estudios Avanzados, Universitat Rovira i Virgili, 2003, pp. 1-10.

Nguz, A., Uncertainty associated with microbiological analysis., Strengthening fishery products health conditions, Belgium, 2007.

Panigrahi, S., Balasubramanian, S., Gu, H., Logue, C., Marchello, M., Design and development of a metal oxide based electronic nose for spoilage classification of beef., Sensors and Actuators, B 119, 2006, pp. 2 - 14.

Perel, M., Kiryanov, A., Krivut, A., Increase in accuracy of the spectrophotometric method of analysis as a result of the introduction of an instrumental correction., Pharmaceutical Chemistry Journal 12, 1978, pp. 1088-1091.

Reutemann, G., Huber, H., Validation of titration methods., Switzerland: Mettler-Toledo, 2003, pp. 1-28.

Walton, R., Validation of Laboratory Tests and Methods., Seminars in Avian and Exotic Pet Medicine ,10, 2001, pp. 59-65.


Standardising and validating aromatic profile analysis by an electronic nose

Carlos Mario Zuluaga1, Amanda Consuelo Díaz2, Martha Cecilia Quicazán3

1 Chemical Engineer, Master in Food Science and Technology, Universidad Nacional de Colombia. Ph.D. student of Chemical Engineering, Universidad Nacional de Colombia, Colombia. cmzuluagad@unal.edu.co

2 Food Engineering , Universidad de la Salle. Doctor Quality, Safety and Food Technology, Universidad de Zaragoza, Spain. Assistant Professor, Universidad Nacional de Colombia. amcdiazmo@unal.edu.co

3 Chemical Engineering, Universidad Nacional de Colombia. Master in Food Science and Technology, Universidad de La Habana, Cuba. Associate Professor, Universidad Nacional de Colombia. mcquicazand@unal.edu.co


ABSTRACT

The electronic nose is one of the most innovative techniques which had been used in the analysis of food and other aroma matrices. This device imitates the human olfactory system through sensors which are partially specific for different groups of volatile chemicals and differentiating and classifying food (by involving multivariate statistics) and it is also used for establishing quality and control parameters. Although the technique has been increasingly applied in research and industry, few studies have been published in the scientific literature referring to validating its use as an analytical methodology. This work' s main objective was to standardise and validate the aromatic profile analysis method for differentiating apiculture products: honey, pollen and propolis. Standardisation involved modifying sample preparation and system variables while validation involved analysis of repeatability and intermediate repeatability of the method and establishing its uncertainty. The results showed that it was possible to set suitable operational parameters for each apicultural product while validation data showed that this technique did not reveal significant differences as reliable values were obtained when using this device as an analysis and quality control tool.

Keywords: Repeatability, intermediate repeatability, uncertainty, aromatic profile, food.


Received: May 10th 2010

Accepted: July 17th 2010

Introduction

There is an ongoing need for developing reliable analytical methods for food analysis which can be evaluated by validation. The Official Journal of the European Community defines validation as, "Confirmation by examination and provision of objective evidence that the particular requirements for a specific intended use have been fulfilled." (CCE, 2002) In other words, it refers to evaluating the performance of an instrument or methodology whose ultimate goal is to provide objective evidence that a particular method will have acceptable and accurate reproducibility to be applicable (Walton, 2001).

The following information should be considered as support for evaluating an analytical procedure: a description of such method including details re operational parameters such as sample preparation, extraction method, reference standards and reagent preparation. The most relevant characteristics when validating an analytical method are specificity, linearity, precision, accuracy, range, detection limits and quantification and robustness which can be partially or fully implemented depending on the nature of the test (APVMA, 2004).

The Universidad Nacional de Colombia' s Institute of Food Science and Technology (ICTA) has been running a research project entitled, "Selecting physical-chemical indicators by applying an electronic nose for cataloguing bee products," (funded by the Ministry of Agriculture and Rural Development). Its objective is to improve apicultural commercial chain competitiveness by developing the scientific and technological tools required for characterising bee products and in the near future to promote quality labels, such as Designation of Origin, for differentiated hive products considering this production chain' s great potential in Colombia.

A methodology for obtaining honey, pollen and propolis' aromatic profile by using an electronic nose was standardised and validated during this project' s first stage; its results are presented in this article. An electronic nose consists of four main elements, it has a technique for extracting aroma or air-flow system which transports volatile compounds to the next step, an array of chemical sensors which transform aroma into electrical signals, an instrumentation and control system for measuring the signal from different sensors and control and automation for the entire system (Mannino et al., 2007). Combining an array of sensors provides a global signal called a "fingerprint" as a simple response (i.e. a characteristic aroma for a particular sample) (López-Feria et al., 2008).

The basic principle of electronic nose operation consists of injecting the headspace (gas space occupied by volatile elements in a vial) with the sample placed in a 20 mL chromatographic vial. Sample amount should not exceed more than 1/3 of vial volume at any time; the sample is then left for an established time and temperature to reach volatile compound concentration equilibrium. A dimensionless conductivity ratio is obtained (G/Go) after injecting the headspace recording time for each electronic nose sensor; this data is related to the conductivity value reported by each sensor in its natural state (Go) and the recorded one when the gaseous substance passes through it (G) (see Figure 1). The expected condition gives a stable response in the shortest time.

The usefulness of electronic noses for classifying and differentiating food industry process control and product quality according to geographical origin has been demonstrated in the scientific literature (Aleixandre et al., 2008; Ampuero et al., 2004; Benedetti et al., 2004; Cosio et al., 2006; Lammertyn et al., 2004); however, no reports were found regarding the methodology' s standardisation and validation for analysing an aromatic profile. The method was thus standardised by finding operating and sample preparation parameters for ensuring an appropriate and reproducible device response. Once such parameters had been found, validation led to evaluating whether the proposed method could be applicable; given the nature of assessing this technique, it was established that the most important characteristics for such validation were accuracy, repeatability and intermediate reproducibility, thereby guaranteeing that the data collected would be reliable and reproducible.

Precision is an instrument' s ability to give the same result when different measurements are made in the same conditions. Repeatability (r) is related to the range of values which may be obtained from an instrument' s response when several measurements are made in the same material, equipment, environmental and operator conditions in a short period of time. It is calculated by subtracting the highest and lowest results of each operator' s analysis and for each electronic nose sensor (equation 1) (INS, 2005).

[1]

Intermediate repeatability (R) is related to the range of values obtained by an instrument' s response within the same laboratory but using different operators, days, equipment and/or reagent lots. Its calculation is similar to that for repeatability but this time by subtracting the lowest from the highest result for all results obtained (Equation 2) (INS, 2005).

[2]

Experimental procedure

Materials

A PEN3 electronic nose (Airsense Analytics GmbH, Germany) was used; this device has a 10 metal oxide sensor (MOS) array (Table 1). Honey, pollen and propolis samples were provided by the beekeepers' associations involved in the aforementioned project.

Standardisation parameters

Standardisation parameters

The same procedure was followed for each hive product, varying both product (sample weight) and equipment operation conditions (vial conditioning temperature, sensor chamber headspace and gas flow stabilisation time) during three stages (Figure 2). The first stage evaluated the sample weight factor with three levels (1, 2 and 3 g), the second assessed two factors (headspace time and stabilisation temperature) having four levels each (5, 10, 15 and 20 min) and the third considered gas flow to the sensor chamber having three levels (60, 100 and 200 mL/min). Three repeats were made for each level. The most appropriate conditions were selected based on the sensors' most stable response and least variability. Once the best value at each stage had been selected it was used as a parameter in the next one.

The coefficient of variation (CV) value was used as factor criteria during standardisation stages; this measurement of dispersion is useful for stating the mean and standard deviation (s), according to equation 3:

[3]

Validation parameters

Honey, pollen and propolis values' repeatability and intermediate repeatability were assessed to validate the analytical methodology with different analysts and on different days. These parameters were established for each electronic nose sensor. Five repeats were made per test. A Student' s t-test was performed for evaluating repeatability to find statistically significant differences between different analysts and different days. An α = 0,05 significance value was used. Microsoft Excel was used for statistical analysis. Equation 4 was used for calculating uncertainty due to precision (U) by relating standard deviation data (SD) and the number of times a sample was analysed (N) (Maroto et al., 2003; Reutemann and Huber, 2003):

[4]

Results

Standardisation

Tables 2, 3 and 4 summarise the results for the honey, pollen and propolis standardisation stage, respectively, showing just the maximum and minimum values from 10 electronic nose sensors' responses. Variation in sample weight (stage I) for honey (Table 2) established that the heavier the sample size, the higher the sensors' coefficient of variation response. It was also observed that these sensors' response stability was greater with a 3 g sample weight so that amount was used in the later stages. Headspace temperature and conditioning time variation revealed that the best response was obtained in 20 min at 40°C; a stable response was given in a shorter analysis time than others, even though not having responses from the sensors having the least variability. This stage also established that a 150 s analysis time was sufficient to ensure response stability. It was found that response was affected by the amount of gas entering the chamber in the last stage; therefore, lower variability was found when dealing with flows as low as 60 mL/min and this would lead to expecting reliable results.

Stage I showed a stable sensor signal in pollen (Table 3)) with a 2 g sample weight. It was found that the amount of sample greatly affected response repeatability. Stage II led to concluding that applying higher temperatures to the vial did not significantly reduce analysis time compared to room temperature; 15 min headspace stabilisation time was enough for obtaining a suitable response, with little variability and in a short injection time (220 s). Unlike the findings for honey analysis, high gas flows into the sensor chamber led to obtaining a more stable response; values greater than 200 mL/min had minimal variability.

Stage I for propolis (Table 4) was conducted with 1 and 2 g weight samples due to the product' s size; higher weights occupied more than 1/3 of the vial. Sample size could not be easily reduced because very small particles may have caused severe damage to the sensors; it was found that 2 g provided stable responses with less variability. Stage II led to observing that heating was unnecessary for obtaining a stable signal response since conductivity values were similar at room temperature; a 15 min stabilisation time offered a less variable response signal. It was determined during this same stage that a 180 s analysis time was sufficient for achieving total stability for the 10 sensors.

Validation

Once the analysis parameters from the standardisation stage had been established then they were validated. The results from calculating the coefficient of variation and repeatability are shown in Table 5 for honey, pollen and propolis; maximum and minimum values found from the 10 electronic nose sensors are shown.

Table 6 shows the range of intermediate repeatability, uncertainty, and t-test values for mean analysis (± 2.7764 critical value).

Overall, it appeared that volatile compound analysis repeatability when using an electronic nose had reduced variability. Some authors have reported less than 2% reference values when assessing the accuracy of methods based on volumetric and chromatographic coefficients of variation (Boqué et al., 2008; Reutemann and Huber, 2003), less than 3% for spectrophotometric methods (Perel et al., 1978) and less than 5% for microbiological methods (Nguz, 2007). Aromatic profile analysis using an electronic nose ranged from 7%-10% for honey, 1% to 4% for pollen and 2%-7% for propolis; this meant that liquid samples has a higher variable response.

T-test analysis results for accuracy between different analysts and different days revealed no statistical differences for all cases, meaning that bee product aromatic profile analysis results using an electronic nose were repeatable for each sensor. Different analysts' uncertainty values were especially low regarding pollen and propolis; uncertainty for honey exceeded 1% in most cases. Sensor response uncertainty was low in pollen and propolis on different days but exceeded 5% in most sensors for honey.

Comparing values with other analytical methods having higher uncertainty, the aromatic profile methodology was reliable, given the specificity and selectivity of this type of analysis compared to chromatographic methods capable of separating compounds, while this approach identified patterns or "fingerprints", groups of substances having common characteristics detected by sensors through changes regarding conductivity. An electronic nose does not work with linear responses, i.e. responses are subject to certain conditions set forth in standardisation allowing more stable responses in the shortest possible analysis time.

Conclusions

A methodology for aromatic profile analysis in honey, pollen and propolis was standardised, the following operating conditions being the most appropriate ones:

Validating analytical methodologies established that methods were statistically precise and repeatable by analysing aromatic profile measurement on different days and by different analysts. Uncertainty values for repeatability and intermediate repeatability for electronic nose sensors in honey ranges from 1% to 10%, 0.5% - 5% in pollen and 0.2% - 5% in propolis meaning thatit can be concluded that honey was the product where sensors provided the most variable response. Liquid samples usually have such pattern while solid matrices tend to have lower uncertainty. These results represent an initial step for future applications of the electronic nose in several fields, especially in quality control. Different laboratories around the world now focus on developing new technologies for assessing substances' aromatic behaviour to find instruments which are able to provide more precise responses. Further investigation is still required in this field.

Acknowledgements

The authors wish to acknowledge the support for carrying out this work provided by the Universidad Nacional de Colombia’s Institute of Food Science and Technology (ICTA) in Bogota, the Ministry of Agriculture and Rural Development, the Sumapaz Region Beekeepers’ Association, the Boyacá Beekeepers’ Association, the Comunera Beekeeping Association, the Association of Beekeepers from Sierra Nevada de Santa Marta, the Colombian Science, Technology and Innovation Department’s (COLCIENCIAS) Young Scientists and Innovative Program and the Science and Food Technology Master’s Program run by the Faculty of Agricultural Sciences at the Universidad Nacional de Colombia in Medellín.


References

Aleixandre, M., Lozano, J., Gutiérrez, J., Sayago, I., Fernández, M., Horrillo, M., Portable e-nose to classify different kinds of wine., Sensors and Actuators, B 131, 2008, pp. 71 - 76.

Ampuero, S., Bogdanov, S., Bosset, J., Classification of unifloral honeys with an MS-based electronic nose, using different sampling modes: SHS, SPME and INDEX., European Journal of Food Research and Technology, 218, 2004, pp. 198-207.

APVMA., Guidelines for the validation of analytical methods for active constituent, agricultural and veterinary chemical products., Australian Pesticides & Veterinary Medicines Authority, Kingston, Australia, 2004.

Benedetti, S., Mannino, S., Sabatini, A., Marcazzan, G., Electronic nose and neural network use for the classification of honey., Apidologie 35, 2004, pp. 397-402.

Boqué, R., Maroto, A., Vander, Y., Assessment of accuracy in chromatographic analysis., LCGC Europe 21, 2008, pp. 15- 21.

CCE., Decisión de la Comisión por la que se aplica la Directiva 96/23/CE del Consejo en cuanto al funcionamiento de los métodos analíticos y la interpretación de los resultados., Diario Oficial de la Unión Europea, 2002.

Cosio, M., Ballabio, D., Benedetti, S., Gigliotti, C., Geographical origin and authentication of extra virgin olive oils by an electronic nose in combination with artificial neural networks., Analytica Chimica Acta, 567, 2006, pp. 202-210.

INS., Selección y validación de métodos de ensayo., Instituto Nacional de Salud. Lima, Perú, 2005.

Lammertyn, J., Veraverbeke, E., Irudayaraj, J., Nose technology for the classification of honey based on rapid aromaprofiling., Sensors and Actuators, B 1, 2004, pp. 54 - 62.

López-Feria, S., Cárdenas, S., García-Mesa, J., Valcárcel, M., Simple and rapid instrumental characterization of sensory attributes of virgin olive oil based on the direct coupling headspace-mass spectrometry., Journal of Chromatography A 1188, 2008, pp. 308 - 313.

Mannino, S., Benedetti, S., Buratti, S., Cosio, M., Scampichio, M., Electrochemical sensors for food authentication., Electrochemical Sensor Analysis, Elsevier, 2007, pp. 1091-1098.

Maroto, A., Boqué, R., Riu, J., Rius, X., Cálculo de incertidumbre en medidas químicas: Análisis cromatográfico., Departamento de Química Analítica y Química Orgánica, Instituto de Estudios Avanzados, Universitat Rovira i Virgili, 2003, pp. 1-10.

Nguz, A., Uncertainty associated with microbiological analysis., Strengthening fishery products health conditions, Belgium, 2007.

Panigrahi, S., Balasubramanian, S., Gu, H., Logue, C., Marchello, M., Design and development of a metal oxide based electronic nose for spoilage classification of beef., Sensors and Actuators, B 119, 2006, pp. 2 - 14.

Perel, M., Kiryanov, A., Krivut, A., Increase in accuracy of the spectrophotometric method of analysis as a result of the introduction of an instrumental correction., Pharmaceutical Chemistry Journal 12, 1978, pp. 1088-1091.

Reutemann, G., Huber, H., Validation of titration methods., Switzerland: Mettler-Toledo, 2003, pp. 1-28.

Walton, R., Validation of Laboratory Tests and Methods., Seminars in Avian and Exotic Pet Medicine ,10, 2001, pp. 59-65.


References

Aleixandre, M., Lozano, J., Gutiérrez, J., Sayago, I., Fernández, M., Horrillo, M., Portable e-nose to classify different kinds of wine., Sensors and Actuators, B 131, 2008, pp. 71 - 76. DOI: https://doi.org/10.1016/j.snb.2007.12.027

Ampuero, S., Bogdanov, S., Bosset, J., Classification of unifloral honeys with an MS-based electronic nose, using different sampling modes: SHS, SPME and INDEX., European Journal of Food Research and Technology, 218, 2004, pp. 198-207. DOI: https://doi.org/10.1007/s00217-003-0834-9

APVMA., Guidelines for the validation of analytical methods for active constituent, agricultural and veterinary chemical products., Australian Pesticides & Veterinary Medicines Authority, Kingston, Australia, 2004.

Benedetti, S., Mannino, S., Sabatini, A., Marcazzan, G., Electronic nose and neural network use for the classification of honey., Apidologie 35, 2004, pp. 397-402. DOI: https://doi.org/10.1051/apido:2004025

Boqué, R., Maroto, A., Vander, Y., Assessment of accuracy in chromatographic analysis., LCGC Europe 21, 2008, pp. 15- 21.

CCE., Decisión de la Comisión por la que se aplica la Directiva 96/23/CE del Consejo en cuanto al funcionamiento de los métodos analíticos y la interpretación de los resultados., Diario Oficial de la Unión Europea, 2002.

Cosio, M., Ballabio, D., Benedetti, S., Gigliotti, C., Geographical origin and authentication of extra virgin olive oils by an electronic nose in combination with artificial neural networks., Analytica Chimica Acta, 567, 2006, pp. 202-210. DOI: https://doi.org/10.1016/j.aca.2006.03.035

INS., Selección y validación de métodos de ensayo., Instituto Nacional de Salud. Lima, Perú, 2005.

Lammertyn, J., Veraverbeke, E., Irudayaraj, J., Nose technology for the classification of honey based on rapid aromaprofiling., Sensors and Actuators, B 1, 2004, pp. 54 - 62. DOI: https://doi.org/10.1016/j.snb.2003.09.012

López-Feria, S., Cárdenas, S., García-Mesa, J., Valcárcel, M., Simple and rapid instrumental characterization of sensory attributes of virgin olive oil based on the direct coupling headspace-mass spectrometry., Journal of Chromatography A 1188, 2008, pp. 308 - 313. DOI: https://doi.org/10.1016/j.chroma.2008.02.046

Mannino, S., Benedetti, S., Buratti, S., Cosio, M., Scampichio, M., Electrochemical sensors for food authentication., Electrochemical Sensor Analysis, Elsevier, 2007, pp. 1091-1098. DOI: https://doi.org/10.1016/S0166-526X(06)49031-0

Maroto, A., Boqué, R., Riu, J., Rius, X., Cálculo de incertidumbre en medidas químicas: Análisis cromatográfico., Departamento de Química Analítica y Química Orgánica, Instituto de Estudios Avanzados, Universitat Rovira i Virgili, 2003, pp. 1-10.

Nguz, A., Uncertainty associated with microbiological analysis., Strengthening fishery products health conditions, Belgium, 2007.

Panigrahi, S., Balasubramanian, S., Gu, H., Logue, C., Marchello, M., Design and development of a metal oxide based electronic nose for spoilage classification of beef., Sensors and Actuators, B 119, 2006, pp. 2 - 14. DOI: https://doi.org/10.1016/j.snb.2005.03.120

Perel, M., Kiryanov, A., Krivut, A., Increase in accuracy of the spectrophotometric method of analysis as a result of the introduction of an instrumental correction., Pharmaceutical Chemistry Journal 12, 1978, pp. 1088-1091. DOI: https://doi.org/10.1007/BF00778075

Reutemann, G., Huber, H., Validation of titration methods., Switzerland: Mettler-Toledo, 2003, pp. 1-28.

Walton, R., Validation of Laboratory Tests and Methods., Seminars in Avian and Exotic Pet Medicine ,10, 2001, pp. 59-65. DOI: https://doi.org/10.1053/saep.2001.22053

How to Cite

APA

Zuluaga, C. M., Díaz, A. C. and Quicazán, M. C. (2011). Standardising and validating aromatic profile analysis by an electronic nose. Ingeniería e Investigación, 31(2), 65–73. https://doi.org/10.15446/ing.investig.v31n2.23466

ACM

[1]
Zuluaga, C.M., Díaz, A.C. and Quicazán, M.C. 2011. Standardising and validating aromatic profile analysis by an electronic nose. Ingeniería e Investigación. 31, 2 (May 2011), 65–73. DOI:https://doi.org/10.15446/ing.investig.v31n2.23466.

ACS

(1)
Zuluaga, C. M.; Díaz, A. C.; Quicazán, M. C. Standardising and validating aromatic profile analysis by an electronic nose. Ing. Inv. 2011, 31, 65-73.

ABNT

ZULUAGA, C. M.; DÍAZ, A. C.; QUICAZÁN, M. C. Standardising and validating aromatic profile analysis by an electronic nose. Ingeniería e Investigación, [S. l.], v. 31, n. 2, p. 65–73, 2011. DOI: 10.15446/ing.investig.v31n2.23466. Disponível em: https://revistas.unal.edu.co/index.php/ingeinv/article/view/23466. Acesso em: 20 apr. 2024.

Chicago

Zuluaga, Carlos Mario, Amanda Consuelo Díaz, and Martha Cecilia Quicazán. 2011. “Standardising and validating aromatic profile analysis by an electronic nose”. Ingeniería E Investigación 31 (2):65-73. https://doi.org/10.15446/ing.investig.v31n2.23466.

Harvard

Zuluaga, C. M., Díaz, A. C. and Quicazán, M. C. (2011) “Standardising and validating aromatic profile analysis by an electronic nose”, Ingeniería e Investigación, 31(2), pp. 65–73. doi: 10.15446/ing.investig.v31n2.23466.

IEEE

[1]
C. M. Zuluaga, A. C. Díaz, and M. C. Quicazán, “Standardising and validating aromatic profile analysis by an electronic nose”, Ing. Inv., vol. 31, no. 2, pp. 65–73, May 2011.

MLA

Zuluaga, C. M., A. C. Díaz, and M. C. Quicazán. “Standardising and validating aromatic profile analysis by an electronic nose”. Ingeniería e Investigación, vol. 31, no. 2, May 2011, pp. 65-73, doi:10.15446/ing.investig.v31n2.23466.

Turabian

Zuluaga, Carlos Mario, Amanda Consuelo Díaz, and Martha Cecilia Quicazán. “Standardising and validating aromatic profile analysis by an electronic nose”. Ingeniería e Investigación 31, no. 2 (May 1, 2011): 65–73. Accessed April 20, 2024. https://revistas.unal.edu.co/index.php/ingeinv/article/view/23466.

Vancouver

1.
Zuluaga CM, Díaz AC, Quicazán MC. Standardising and validating aromatic profile analysis by an electronic nose. Ing. Inv. [Internet]. 2011 May 1 [cited 2024 Apr. 20];31(2):65-73. Available from: https://revistas.unal.edu.co/index.php/ingeinv/article/view/23466

Download Citation

CrossRef Cited-by

CrossRef citations4

1. Carlos Mario Zuluaga-Domínguez, Amanda Consuelo Díaz-Moreno, Carlos Alberto Fuenmayor, Martha Cecilia Quicazán. (2013). Pot-Honey. , p.417. https://doi.org/10.1007/978-1-4614-4960-7_30.

2. Carlos Mario Zuluaga-Domínguez, Andrea Nieto-Veloza, Marta Quicazán-de-Cuenca. (2018). Classification of Colombian honeys by electronic nose and physical-chemical parameters, using neural networks and genetic algorithms. Journal of Apicultural Research, 57(1), p.145. https://doi.org/10.1080/00218839.2017.1339521.

3. Carlos Mario Zuluaga‐Domínguez, Carlos Alberto Fuenmayor, Marta Cecilia Quicazán. (2023). Bioactive Attributes and Analysis of Electronic Nose Feature Signals of Colombian Stingless Bees Propolis. Chemistry & Biodiversity, 20(1) https://doi.org/10.1002/cbdv.202200952.

4. Luisa Juana Bernal, Laura Angélica Melo, Consuelo Díaz Moreno. (2014). Evaluation of the Antioxidant Properties and Aromatic Profile During Maturation of The Blackberry (Rubus glaucus Benth) and The Bilberry (Vaccinium meridionale Swartz). Revista Facultad Nacional de Agronomía Medellín, 67(1), p.7209. https://doi.org/10.15446/rfnam.v67n1.42649.

Dimensions

PlumX

Article abstract page views

2471

Downloads

Download data is not yet available.