Final prototype of the concentrator. (a) Front View, (b) Side View

Publicado

2025-05-16

Design of an oxygen concentrator according to ISO 80601-2-69 standards

Diseño de un concentrador de oxígeno de acuerdo con las normas ISO 80601-2-69

DOI:

https://doi.org/10.15446/dyna.v92n237.117197

Palabras clave:

oxygen, concentrator, design, feasibility, economics (en)
oxígeno, concentrador, diseño, factibilidad, economía (es)

Descargas

Autores/as

This work focuses on the design, construction and economic analysis of an oxygen concentrator in the city of Moquegua, Peru, under ISO 80601-2-69 standards. The methodology used was experimental, consisting of the design and construction of an oxygen concentrator, as well as its economic analysis. The results of the prototype tests indicate that it provides oxygen with a concentration of 95%, up to 10 litres/min, consuming 280 W of power, being powered by 220 V alternating current, working under a process called oscillating pressure absorption. The final cost of the device was S/1220.80, which contrasts with the cost of similar equipment that in the market have a value of S/1500.00 and that in pandemic 2020 reached S/15000.00. In conclusion, the manufacture of an oxygen concentrator in Moquegua is both technically and economically feasible.

Este trabajo se enfoca en el diseño, construcción y análisis económico de un concentrador de oxígeno en la ciudad de Moquegua, Perú, bajo las normas ISO 80601-2-69. La metodología utilizada fue experimental, que consistió en el diseño y construcción de un concentrador de oxígeno, así como el análisis económico del mismo. Los resultados de las pruebas del prototipo indican que proporciona oxígeno con una concentración del 95%, hasta por 10 litros/min, consumiendo 280 W de potencia, siendo alimentado por corriente alterna de 220 V, trabajando bajo un proceso denominado absorción por presión oscilante. El costo final del dispositivo fue de S/1220.80, que contrasta con el costo de equipos similares que en el mercado tienen un valor de S/1500.00 y que en pandemia del 2020 alcanzó los S/15000.00. En conclusión, la fabricación de un concentrador de oxígeno en Moquegua resulta factible tanto técnica como económicamente.

Referencias

[1] Los, E., Ford, G., Tuell, D., Macariola, D., and Stone, W., The roles of glutathione and oxidative stress in diabetes and COVID-19. Oxygen, 4(3), pp. 351–362, 2024. DOI: https://doi.org/10.3390/oxygen4030021

[2] Dargent, E. and Gianella, C., País sin Oxígeno: debilidad institucional en COVID-19. Revista de Ciencia Política (Santiago), 44(1), pp. 87–107, 2024. DOI: https://doi.org/10.4067/s0718-090x2024005000104

[3] Huo, M., Wang, L., Chen, Y., and Shi, J., Oxygen pathology and oxygen-functional materials for therapeutics. Matter, 2(5), pp. 1115–1147, 2020. DOI: https://doi.org/10.1016/j.matt.2020.02.013

[4] Van Vliet, T., Casciaro, F., and Demaria, M., To breathe or not to breathe: understanding how oxygen sensing contributes to age-related phenotypes. Ageing Res Rev, 67, art. 101267, 2021. DOI: https://doi.org/10.1016/j.arr.2021.101267

[5] Hocke, K., Oxygen in the earth system. Oxygen, 3(3), pp. 287–299, 2023. DOI: https://doi.org/10.3390/oxygen3030019

[6] Yunus, B., Sri, A., and Teguh, K., Design and construction of portable oxygen concentrator using pressure swing adsorption technology. ASEAN Journal for Science and Engineering in Materials, [online]. 3(1), pp. 43–50, 2024. Available: https://ejournal.bumipublikasinusantara.id/index.php/ajsem/article/view/413

[7] West, J., The strange history of atmospheric oxygen. Physiol Rep, 10(6), 2022. DOI: https://doi.org/10.14814/phy2.15214

[8] Shi, P., et al. Factors contributing to spatial–temporal variations of observed oxygen concentration over the Qinghai-Tibetan Plateau. Sci Rep, 11(1), 2021. DOI: https://doi.org/10.1038/s41598-021-96741-6

[9] Hancock, J., A brief history of oxygen: 250 years on. Oxygen, 2(1), pp. 31–39, 2022. DOI: https://doi.org/10.3390/oxygen2010004

[10] Zou, Q., Lai, Y., and Lun, Z.-R., Exploring the association between oxygen concentration and life expectancy in China: a quantitative analysis. Int J Environ Res Public Health, 20(2), art. 1125, Jan. 2023. DOI: https://doi.org/10.3390/ijerph20021125

[11] Madankar, T.A., and Deshpande, Y., Development of portable oxygen concentrator - A review. E3S Web of Conferences, 430, 2023. DOI: https://doi.org/10.1051/e3sconf/202343001265

[12] López, J., Escarramán, D., Sánchez, J., and Pérez, O., Ratio de oxígeno como determinante de severidad en neumonía COVID-19. Revista Chilena de Anestesia, 52(7), pp. 678–682, 2023. DOI: https://doi.org/10.25237/revchilanestv52n7-08

[13] Martínez, E., Morales, G., Segura, L., and De la Cruz, A. Métodos no invasivos de oxigenación en pacientes con COVID-19. Revisión descriptiva. Medicina Crítica, 36(6), pp. 378–386, 2022. DOI: https://doi.org/10.35366/107461

[14] Rafathnazneen, N., Vijayalaxmi, H., Tamaloon, S., Varun, S., and Jagadish, J., Development of low cost and portable oxygen concentrator. International Journal of Research Publication and Reviews, [online]. 4(5), pp. 1070–1074, 2023. Available at: https://ijrpr.com/uploads/V4ISSUE5/IJRPR12779.pdf

[15] Lustenberger, U., Krestnikova, A., Gröninger, O., Grass, R., and Stark, W., Knowledge transfer in support of the development of oxygen concentrators in emergency settings during the COVID-19 Pandemic. J Chem Educ, 100(5), pp. 1858–1865, 2023. DOI: https://doi.org/10.1021/acs.jchemed.2c00925

[16] Zhang, Q., et al., Experimental study on oxygen concentrator with wide product flow rate range: individual parametric effect and process improvement strategy. Sep Purif Technol, 274, 2021. DOI: https://doi.org/10.1016/j.seppur.2021.118918

[17] Arora and Hasan, F., Flexible oxygen concentrators for medical applications. Sci Rep, 11, 14317, 2021. DOI: https://doi.org/10.1038/s41598-021-93796-3

[18] Gadiraju, N., et al., Design and development of a novel system for remote control of stationary oxygen concentrator flow rate. Medical Devices: Evidence and Research, 16, pp. 91–100, 2023. http://doi.org/10.2147/MDER.S407233

Cómo citar

IEEE

[1]
V. W. Mamani-Cori, N. J. Zeballos-Hurtado, y A. J. Cosi-Blancas, «Design of an oxygen concentrator according to ISO 80601-2-69 standards», DYNA, vol. 92, n.º 237, pp. 59–64, may 2025.

ACM

[1]
Mamani-Cori , V.W., Zeballos-Hurtado, N.J. y Cosi-Blancas, A.J. 2025. Design of an oxygen concentrator according to ISO 80601-2-69 standards. DYNA. 92, 237 (may 2025), 59–64. DOI:https://doi.org/10.15446/dyna.v92n237.117197.

ACS

(1)
Mamani-Cori , V. W.; Zeballos-Hurtado, N. J.; Cosi-Blancas, A. J. Design of an oxygen concentrator according to ISO 80601-2-69 standards. DYNA 2025, 92, 59-64.

APA

Mamani-Cori , V. W., Zeballos-Hurtado, N. J. & Cosi-Blancas, A. J. (2025). Design of an oxygen concentrator according to ISO 80601-2-69 standards. DYNA, 92(237), 59–64. https://doi.org/10.15446/dyna.v92n237.117197

ABNT

MAMANI-CORI , V. W.; ZEBALLOS-HURTADO, N. J.; COSI-BLANCAS, A. J. Design of an oxygen concentrator according to ISO 80601-2-69 standards. DYNA, [S. l.], v. 92, n. 237, p. 59–64, 2025. DOI: 10.15446/dyna.v92n237.117197. Disponível em: https://revistas.unal.edu.co/index.php/dyna/article/view/117197. Acesso em: 26 dic. 2025.

Chicago

Mamani-Cori , Vigil Wuilber, Nilton Juan Zeballos-Hurtado, y Arturo Jesús Cosi-Blancas. 2025. «Design of an oxygen concentrator according to ISO 80601-2-69 standards». DYNA 92 (237):59-64. https://doi.org/10.15446/dyna.v92n237.117197.

Harvard

Mamani-Cori , V. W., Zeballos-Hurtado, N. J. y Cosi-Blancas, A. J. (2025) «Design of an oxygen concentrator according to ISO 80601-2-69 standards», DYNA, 92(237), pp. 59–64. doi: 10.15446/dyna.v92n237.117197.

MLA

Mamani-Cori , V. W., N. J. Zeballos-Hurtado, y A. J. Cosi-Blancas. «Design of an oxygen concentrator according to ISO 80601-2-69 standards». DYNA, vol. 92, n.º 237, mayo de 2025, pp. 59-64, doi:10.15446/dyna.v92n237.117197.

Turabian

Mamani-Cori , Vigil Wuilber, Nilton Juan Zeballos-Hurtado, y Arturo Jesús Cosi-Blancas. «Design of an oxygen concentrator according to ISO 80601-2-69 standards». DYNA 92, no. 237 (mayo 9, 2025): 59–64. Accedido diciembre 26, 2025. https://revistas.unal.edu.co/index.php/dyna/article/view/117197.

Vancouver

1.
Mamani-Cori VW, Zeballos-Hurtado NJ, Cosi-Blancas AJ. Design of an oxygen concentrator according to ISO 80601-2-69 standards. DYNA [Internet]. 9 de mayo de 2025 [citado 26 de diciembre de 2025];92(237):59-64. Disponible en: https://revistas.unal.edu.co/index.php/dyna/article/view/117197

Descargar cita

CrossRef Cited-by

CrossRef citations0

Dimensions

PlumX

Visitas a la página del resumen del artículo

211

Descargas

Los datos de descargas todavía no están disponibles.