Design and validation of a prototype to calibrate the pressure variable in tonometry

Diseño y validación de un prototipo para calibrar la variable presión en tonometría

DOI:

https://doi.org/10.15446/dyna.v88n218.91715

Palabras clave:

tonometry, metrology, uncertainty in measurement, calibration (en)
tonometría, metrología, incertidumbre en la medición, calibración (es)

Autores/as

This paper corresponds to the results of an experimental research that to led to the design of a useful prototype for quality control of measurements in the field of tonometry to diagnose and check glaucoma. All tonometry techniques are based on Goldmann ́s tonometers (GT) and the Imbert-Fick principle. Taking into account the principles and techniques of industrial metrology, the prototype could be used as a standard to calibrate the pressure readings of this class of tonometers. The developed protoype eliminates the indirect measurement that is required with current standards and facilitates a greater control of the sources of error in the measurement. The model is based on the integration of a sensor to a hydraulic model to measure pressures from 0 to 80 mmHg range, whose precision was validated through R&R studies. The precision of the prototype was validated according to the Rigel UNI-SIM standard and was complemented by calibrating 15 tonometers.

Este artículo corresponde a los resultados de una investigación experimental que derivó en el diseño de un prototipo útil para el control de calidad de las mediciones en el campo de la tonometría para diagnosticar y controlar el glaucoma. Todas las técnicas de tonometría se basan en los tonómetros de Goldmann (TG) y en el principio de Imbert-Fick. Teniendo en cuenta los principios y técnicas de la metrología industrial, el prototipo podría usarse como un estándar para calibrar las lecturas de presión de esta clase de tonómetros. El prototipo desarrollado elimina la medición indirecta que se requiere con los patrones actuales y facilita un control mayor de las fuentes de error en la medición. El modelo se basa en la integración de un sensor a un modelo hidráulico para medir presiones de 0 a 80 mmHg de alcance, cuya precisión se validó a través de experimentos R&R. La precisión del prototipo fue validada según el estándar Rigel UNI-SIM y se complementó mediante la calibración de 15 tonómetros.

Referencias

Goldmann, H. (1957). Uber Applanations tonometrie. In Classic Papers in Glaucoma (p. 153).

Stuckey, G. C. (1974). Tonometry and the Imbert‐Fick Law. Australian Journal of Opthalmology, 2(2), 68-70.

Cook, J. A., Botello, A. P., Elders, A., Ali, A. F., Azuara-Blanco, A., Fraser, C., ... & Surveillance of Ocular Hypertension Study Group. (2012). Systematic review of the agreement of tonometers with Goldmann applanation tonometry. Ophthalmology, 119(8), 1552-1557.

Stamper RL. A history of intraocular pressure and its measurement. Optom Vis Sci. 2011;88:E16–28

ISO/IEC 17025:2017. General Requirements for the Competence of Testing and Calibration Laboratories; International Organization for Standardization: Geneva, Switzerland, 2017.

Tobón, O., & Rodríguez, V. (2017). Desarrollo y estandarización de métodos de calibración para equipos utilizados en salud visual (Queratómetros, Lensómetros y Tonómetros), implementados en el Hospital Universitario de San Vicente Fundación. Revista Ingeniería Biomédica, 11(22), 13–20. https://doi.org/10.24050/19099762.n22.2017.1179

Choudhari, N. S., George, R., & Vijaya, L. (2008). Have You Tested Your Goldmann Tonometer Today? Journal of Current Glaucoma Practice, 2(2), 13–16. Disponible en: https://doi.org/10.5005/jp-journals-10008-1027.

Kirstein, E. M., Elsheikh, A., & Gunvant, P. (2011). Tonometry–past, present and future. Glaucoma–Current Clinical and Research Aspects. InTech, 85-108.

Cho, P., & Lui, T. (1997). Comparison of the performance of the Nidek NT-2000 noncontact tonometer with the Keeler Pulsair 2000 and the Goldmann applanation tonometer. Optometry and vision science: official publication of the American Academy of Optometry, 74(1), 51-58.

Moseley, M. J., Thompson, J. R., Deutsch, J., Misson, G. P., Naylor, G., Tan-Yee, A., ... & Fielder, A. R. (1993). Comparison of the Keeler Pulsair 2000 non-contact tonometer with Goldmann applanation. Eye, 7(1), 127-130.

Leonardi, M., Leuenberger, P., Bertrand, D., Bertsch, A., & Renaud, P. (2004). First Steps toward Noninvasive Intraocular Pressure Monitoring with a Sensing Contact Lens. Investigative Opthalmology & Visual Science, 45(9), 3113–3117. Disponible en: https://doi.org/10.1167/iovs.04-0015

McCafferty, S., Lim, G., Duncan, W., Enikov, E., & Schwiegerling, J. (2016). Goldmann Tonometer Prism with an Optimized Error Correcting Applanation Surface. Translational Vision Science & Technology, 5(5), 4. Dsiponible en: https://doi.org/10.1167/tvst.5.5.4

Bowling, B. (2015). Cornea. In B. Bowling, Kanski's Clinical Ophthalmology (pp. 167-237). Elsevier.

Stapleton, F., Stretton, S., Papas, E., Skotnitsky, C., & Sweeney, D. F. (2006). Silicone hydrogel contact lenses and the ocular surface. The ocular surface, 4(1), 24-43.

Villate, M. A. M., Méndez, J. D. B., & Echeverry, J. E. P. (2016). Anatomía quirúrgica del ojo: Revisión anatómica del ojo humano y comparación con el ojo porcino. Morfolia, 8(3), 21-44.

Montgomery, D. C. (2007). Introduction to statistical quality control. 6 th Edition. John Wiley & Sons, pp.368-383.

Pennella, R. (2004). Managing the metrology system. Milwaukee, Wis.: ASQ Quality Press, Third edition. 212 p.

Higuchi, M., Hotta, M., & Iwabuchi, M. (1992). U.S. Patent No. 5,121,119. Washington, DC: U.S. Patent and Trademark Office.

International Organization for Standardization ISO. (2009). International Standard ISO 8612 Ophthalmic instruments — Tonometers. International Organization for Standardization. Disponible en: https://doi.org/10.1109/IEEESTD.2010.5733835

Alm, A., & Kaufman, P. L. (Eds.). (2003). Adler's Physiology of the Eye: Clinical Application, 10th Edition, Mosby: St. Louis. 876 pp.

Fannin, T. E., & Grosvenor, T. (2013). Clinical optics. Butterworth-Heinemann. 466 pp.

Andreanos, K., Koutsandrea, C., Papaconstantinou, D., Diagourtas, A., Kotoulas, A., Dimitrakas, P., & Moschos, M. M. (2016). Comparison of goldmann applanation tonometry and pascal dynamic contour tonometry in relation to central corneal thickness and corneal curvature. Clinical Ophthalmology, 10, 2477–2484. https://doi.org/10.2147/OPTH.S115203

McAlinden, C., Khadka, J., & Pesudovs, K. (2011). Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmology. Ophthalmic and Physiological Optics, 31(4), 330-338.

Mansouri, K., Tanna, A. P., De Moraes, C. G., Camp, A. S., & Weinreb, R. N. (2020). Review of the measurement and management of 24-hour intraocular pressure in patients with glaucoma. Survey of ophthalmology, 65(2), 171-186. http://dx.doi.org/10.1016/j.survophthal.2019.09.004

Pascolini, D., & Mariotti, S. P. (2012). Global estimates of visual impairment: 2010. British Journal of Ophthalmology, 96(5), 614-618.

Sródka, W. (2011). Evaluating the material parameters of the human cornea in a numerical model. Acta of Bioengineering and Biomechanics / Wrocław University of Technology, 13(3), 77–85.

OTHER CITED MATERIAL

A. Haag-Streit (2017). Instructions for use Applanation tonometer AT 900® / AT 870. Koeniz. Acces. Available at https://www.haag-streit.com/fileadmin/Haag-Streit_USA/Diagnostics/tonometry/download/01-IFU_AT900-7006000-04230_eng_01.pdf. Accesed January 18, 2019.B.

Appasamy Associates. Empowerming Vision. Available at https://www.appasamy.com/equipmenthome.phpproduct=Appa_applanation_tonometer_AATM_5001.C.

Gomez, E. (2017). Ajustar señal para ADC y aprovecha el ADC completo. Available at. https://www.rinconingenieril.es/ajustar-senal-adc/. Accesed Febrero 05, 2019. [D] Inopto. (2016). T10 Applanation Tonometer User Manual.

Cómo citar

IEEE

[1]
U. F. Pineda Zapata y J. F. . Daza-Fernández, «Design and validation of a prototype to calibrate the pressure variable in tonometry», DYNA, vol. 88, n.º 218, pp. 185–193, jul. 2021.

ACM

[1]
Pineda Zapata, U.F. y Daza-Fernández, J.F. 2021. Design and validation of a prototype to calibrate the pressure variable in tonometry. DYNA. 88, 218 (jul. 2021), 185–193. DOI:https://doi.org/10.15446/dyna.v88n218.91715.

ACS

(1)
Pineda Zapata, U. F.; Daza-Fernández, J. F. . Design and validation of a prototype to calibrate the pressure variable in tonometry. DYNA 2021, 88, 185-193.

APA

Pineda Zapata, U. F. & Daza-Fernández, J. F. . (2021). Design and validation of a prototype to calibrate the pressure variable in tonometry. DYNA, 88(218), 185–193. https://doi.org/10.15446/dyna.v88n218.91715

ABNT

PINEDA ZAPATA, U. F.; DAZA-FERNÁNDEZ, J. F. . Design and validation of a prototype to calibrate the pressure variable in tonometry. DYNA, [S. l.], v. 88, n. 218, p. 185–193, 2021. DOI: 10.15446/dyna.v88n218.91715. Disponível em: https://revistas.unal.edu.co/index.php/dyna/article/view/91715. Acesso em: 7 mar. 2026.

Chicago

Pineda Zapata, Uriel Fernando, y Juan Fernando Daza-Fernández. 2021. «Design and validation of a prototype to calibrate the pressure variable in tonometry». DYNA 88 (218):185-93. https://doi.org/10.15446/dyna.v88n218.91715.

Harvard

Pineda Zapata, U. F. y Daza-Fernández, J. F. . (2021) «Design and validation of a prototype to calibrate the pressure variable in tonometry», DYNA, 88(218), pp. 185–193. doi: 10.15446/dyna.v88n218.91715.

MLA

Pineda Zapata, U. F., y J. F. . Daza-Fernández. «Design and validation of a prototype to calibrate the pressure variable in tonometry». DYNA, vol. 88, n.º 218, julio de 2021, pp. 185-93, doi:10.15446/dyna.v88n218.91715.

Turabian

Pineda Zapata, Uriel Fernando, y Juan Fernando Daza-Fernández. «Design and validation of a prototype to calibrate the pressure variable in tonometry». DYNA 88, no. 218 (julio 28, 2021): 185–193. Accedido marzo 7, 2026. https://revistas.unal.edu.co/index.php/dyna/article/view/91715.

Vancouver

1.
Pineda Zapata UF, Daza-Fernández JF. Design and validation of a prototype to calibrate the pressure variable in tonometry. DYNA [Internet]. 28 de julio de 2021 [citado 7 de marzo de 2026];88(218):185-93. Disponible en: https://revistas.unal.edu.co/index.php/dyna/article/view/91715

Descargar cita

CrossRef Cited-by

CrossRef citations0

Dimensions

PlumX

Visitas a la página del resumen del artículo

463

Descargas

Los datos de descargas todavía no están disponibles.