Published

2014-01-01

Automixer: equipment for the reduction of risks associated with inadequate oxygen supply

Automixer: equipo para disminuir los riesgos por el inadecuado suministro de oxígeno

DOI:

https://doi.org/10.15446/ing.investig.v34n1.42767

Keywords:

Blender, fuzzy logic, newborn, pulse oxymetry, oxygen saturation (en)
mezclador, lógica difusa, neonato, pulsoximetría, saturación de oxígeno (es)

Downloads

Authors

  • Jesús A. López Universidad Autónoma de Occidente
  • Robinson Araque Campo Atem Health Care Research and development center
  • Alejandro Matiz Rubio Atem Health Care Research and development center

In this paper, the AUTOMIXER, a medical support equipment for the reduction of risks associated with the inadequate supply of oxygen, is presented. The equipment automatically regulates the Inspired Oxygen Fraction (FIO2 concentration), which is supplied to the patient according to user parameters, measures of oxygen saturation and patient's status. The AUTOMIXER is based on a fuzzy control system, which was implemented in a microcontroller using a LookUp Table. The operation of the device was validated using simulated data and actual patients. The results show that the AUTOMIXER is an important tool for decreasing the incidence of pathologies related to unsuitable handling (excess or lack) of oxygen.

En este trabajo se presenta un equipo de apoyo médico, para disminuir el riesgo por suministro inadecuado de oxígeno; denominado AUTOMIXER. El equipo desarrollado regula de manera automática y precisa la Fracción Inspirada de Oxígeno (concentraciónFiO2), que se le suministra al paciente de acuerdo a los parámetros programados por el usuario, las medidas de la saturación de oxígeno y el estado del paciente. El AUTOMIXER está basado en un sistema de control difuso, el cual fue implementado en un micro controlador usando una LookUp Table. Para validar el funcionamiento del dispositivo, se realizaron diferentes pruebas con datos simulados y con pacientes reales; los resultados muestran que el AUTOMIXER se vislumbra como una importante herramienta para disminuir la incidencia de las patologías, asociadas al manejo inadecuado (exceso o falta) del oxígeno.

References

Azad, V., Retinopathy of Prematurity: A Text and Atlas., Jaypee Brothers Medical Publishers, 2006.

Caicedo, E., López, J., Muñoz, A., Control Inteligente., EBOOK, Colombia, Editorial Universidad del Valle, 2012.

Claure, N., D'Ugard, C., Bancalari, E., Multicenter crossover study of automated control of inspired oxygen in ventilated preterm infants., Pediatrics, Vol. 127, No. 1, 2011.

Donoghue, D., Bawden, K., Cartwright, D., Darlow, B., Henderson Smart, D., Lancaster, P., The report of the Australian and New Zealand Neonatal Network, 2000., Sydney, 2002.

Kosko, B., Neural Networks and Fuzzy Systems: A Dynamical Systems Approach to Machine Intelligence., Prentice Hall, 1991.

Matiz, A., Araque, R., Zapata, J., Mezclador automático de autoregulación de oxígeno con monitoreo no invasivo por pulsoximetría., PCT/ID2007/002157, 27 Julio, 2007.

Ogata, K., Ingeniería de Control Moderna, 2nd ed., México, Prentice Hall, 1993.

Passino, K., Yurkovich, S., Fuzzy Control, Addison Wesley, 1998.

Smith, M., Rehabilitación cardiovascular y respiratoria., España, Harcourt Brace, 2000.

Strickland, Jr., James, H., Medical ventilator system., 2004.

West, J., Pulmonary Physiology and Pathophysiology: An Integrated, Case Based Approach., 1st ed., Lippincott Williams & Wilkins, 2001.

Zin, A., De la Fuente, M., Gilbert, C., Quinn, G., Sola, A., Ophthalmic and neonatal guidelines for ROP screening and treatment in Latin American countries., October 2007.