Publicado

2019-01-01

Técnicas de titulación ácido-base: consideraciones metrológicas

Acid-base titration techiniques: metrological considerations

Técnicas titulação ácido-base: considerações metrológicas

DOI:

https://doi.org/10.15446/rev.colomb.quim.v48n1.72401

Palabras clave:

Incertidumbre de medición, punto de equivalencia, indicador ácido-base, potenciometría, conductividad electrolítica, gravimetría, volumetría (es)
Uncertainty measurement, equivalence point, acid-base indicator, potentiometry, electrolytic conductivity, gravimetric analysis, volumetric analysis (en)
Incerteza de medição, análise gravimétrica, análise volumétrica (pt)

Autores/as

  • Diego Alejandro Ahumada Forigua Subdirección de metrología Química y Biomedicina. Instituto Nacional de Metrología de Colombia
  • Laura Vanessa Morales Erazo Subdirección de metrología Química. Instituto Nacional de Metrología de Colombia
  • Johanna Paola Abella Gamba Subdirección de metrología Química. Instituto Nacional de Metrología de Colombia
  • Ivonne Alejandra Gonzalez Cardenas Subdirección de metrología Química y Biomedicina. Instituto Nacional de Metrología de Colombia

Se evaluaron diferentes sistemas de medición para titulaciones ácido-base, con el propósito de determinar cuál presenta mejores características metrológicas. Estos sistemas incluyen volumetrías y gravimetrías, las cuales se combinaron con diferentes métodos de detección del punto de equivalencia: (i) potenciometría, (ii) indicador ácido-base y (iii) conductividad electrolítica. Este artículo propone una nueva aproximación para la estimación de la incertidumbre de medición asociada a la detección del punto de equivalencia. Los resultados indicaron que, para la determinación del punto de equivalencia, las incertidumbres más bajas se obtienen mediante el método potenciométrico en comparación con el método conductimétrico cuyo sesgo es hasta 78 veces mayor. Sin embargo, se encontró que el método potenciométrico es fuertemente dependiente de la resolución del instrumento y/o la cantidad de agente titulante que se adiciona en la cercanía al punto de equivalencia. También se encontró que la incertidumbre disminuye cerca del 40% empleando métodos gravimétricos en comparación con los volumétricos. Al realizar la evaluación del sesgo de los sistemas de medición acorde con la guía ISO 33:2015, los resultados mostraron que, de los métodos evaluados, el método gravimétrico con detección potenciométrica presentó las mejores características metrológicas, pues se obtuvieron sesgos e incertidumbres menores.

Different measurement systems for acid-base titrations were evaluated to determine which one had the best metrological characteristics. These systems included volumetric and gravimetric titrations, which were combined with different detection systems: (i) potentiometry, (ii) acid-base indicator, and (iii) electrolytic conductivity. This article proposes a new approach for estimating the measurement uncertainty associated with the detection of the equivalence point. The results showed that, for the determination of the equivalence point, the lowest uncertainties were obtained by the potentiometric method in comparison with the conductometric method, whose bias was 78 times higher. The potentiometric method, however, was strongly dependent on the instrument resolution and/or the titrant amount that is added close to the equivalence point. It was also found that uncertainty decreases about 40% using gravimetric methods as compared to volumetric methods. When performing bias evaluation of the measurement systems according to the ISO 33:2015 guide, results showed that, among the evaluated methods, the gravimetric method with potentiometric detection presented the best metrological characteristics, with the lowest biases and uncertainties.

Foram avaliados diferentes sistemas de medida para titulações ácido-base, a fim de determinar quais possuem as melhores características metrológicas. Estes sistemas incluem titulações e gravimétricas, que foram combinadas com diferentes sistemas de detecção do ponto de equivalência: (i) potenciometria, (ii) indicador ácido-base e (iii) condutividade electrolítica. Este artigo propõe uma nova abordagem para estimar a incerteza de medição associada à detecção do ponto de equivalência. Os resultados indicaram que, para a determinação do ponto de equivalência, as incertezas mais baixas são obtidas pelo método potenciométrico enquanto que a incerteza do método condutométrico é até 78 vezes maior que a obtida pela potenciometria. Além disso, a potencimetria  é altamente dependente da resolução do instrumento e/ou da quantidade de titulante para ser adicionado na proximidade com o ponto de equivalência. Também verificou-se que a incerteza diminui em torno de 40% utilizando métodos gravimétricos. De acordo à guia de norma ISO 33:2015, os resultados mostraram que os métodos de titulação ácido-base avaliados gravimetricamente com detecção potenciométrica proporcionaram as melhores características de metrologia, porque obtiveram os menores distorções e incertezas.

Referencias

A. Johansson. “The Development of the Titration Methods. Some Historical Annotations” Anal. Chim. Acta, vol. 206, pp. 97–109, 1988. DOI: http://doi.org/10.1016/S0003-2670(00)80834-X

R. Madsen. The Development of Titrimetric Analysis till 1806. Copenhagen: Gad Publishers, 2016, p. 19

E. Koort, K. Herodes, V. Pihl, I. Leito. “Estimation of Uncertainty in pKa values determined by potentiometric titration” Anal. Bioanal. Chem, vol. 379, pp. 720–729, 2004. DOI: http://doi.org/10.1007/ s00216-004-2586-1

K. Skic, P. Boguta, Z. Sokołowska. “Analysis of the sorption properties of different soils using water vapour adsorption and potentiometric titration methods” Int. Agrophys, vol. 30, pp. 369–374, 2016. DOI: https://doi.org/10.1515/intag-2015-0100

J. Lochhead, M. Nelson, A. Schneider. “Risks and Benefits of Rapid Clozapine Titration” Ment. Illn, vol. 8, pp. 6457-6461, 2016. DOI: https://doi.org/10.4081/mi.2016.6457

S. Quintar, J. Santagata, V. Cortinez. “Determination of vanadium(V) by direct automatic potentiometric titration with EDTA using a chemically modified electrode as a potentiometric sensor” Talanta, vol. 67, pp. 843–847, 2005. DOI: https://doi.org/10.1016/j.talanta.2005.04.010

H. Wang, N. Pampati, W. McCormick, L. Bhattacharyya. “Protein Nitrogen Determination by Kjeldahl Digestion and Ion Chromatography” J. Pharm. Sci, vol. 105, pp. 1851–1857, 2016. DOI: https://doi.org/10.1016/j.xphs.2016.03.039

R. Klet, Y. Liu, T.Wang, J. Hupp, O. Farha. “Evaluation of Brønsted acidity and proton topology in Zr- and Hf-based metal–organic frameworks using potentiometric acid–base titration” J. Mater. Chem. A, vol. 4, pp. 1479-1485, 2016. DOI: https://doi.org/10.1039/ C5TA07687K

C. Ophardt. “Acid Rain Analysis by Standard Addition Titration”. J. Chem. Educ, vol. 62, pp. 257–258, 1985. DOI: https://doi.org/10.1021/ ed062p257

K. Pratt. “Automated, High-Precision Coulometric Titrimetry Part II. Strong and Weak Acids and Bases” Anal. Chim. Acta, vol. 289, pp. 135–142, 1994. DOI: https://doi.org/10.1016/0003-2670(94)80096-0

I. Kuselman, A. Shenhar. “Uncertainty in chemical analysis and validation of the analytical method: acid value determination in oils” Accredit. Qual. Assur, vol. 2 pp. 180-185, 1992. DOI: https://doi. org/10.1007/s007690050127

R. Capita, S. Llorente-Marigómez, M. Prieto, C. Alonso-Calleja. “Microbiological Profiles, pH, and Titratable Acidity of Chorizo and Salchichón (Two Spanish Dry Fermented Sausages) Manufactured with Ostrich, Deer, or Pork Meat” J. Food Prot, vol. 69, pp. 1183–1189, 2006. DOI: https://doi.org/10.4315/0362-028X-69.5.1183

M. Milton, T. Quinn. “Primary methods for the measurement of amount of substance” Metrologia., vol. 38, pp. 289–296, 2001. DOI: https://doi. org/10.1088/0026-1394/38/4/1

H. Felber, S. Rezzonico, M. Máriássy. “Titrimetry at a metrological level” Metrologia, vol. 40, 249–254, 2006. DOI: https://doi. org/10.1088/0026-1394/40/5/006

Instituto Nacional de Metrología. “Evaluación de los resultados de la encuesta de demanda metrológica en la Industria Colombiana” [En línea]. 2011. Disponible: http://www.inm.gov.co/images/contenidos/ EstudiosSector/INFORME%20EDMIC%20I.pdf. [Último acceso: 31 07 2018]

Chavarro, L. A.; Porras, C. E. “Introducción Al Desarrollo de Capacidad Para Producir, Analizar Y Evaluar Materiales de Referencia” 2009. [En línea]. Disponible: http://www.inm.gov.co/images/ contenidos/EstudiosSector/Estudio_diagnostico_de_las_necesidades_ en_metrologia_en_quimica_en_tres_sectores_estrategicos_de_la_ economia_de_Colombia.pdf. [Último acceso: 31 07 2018]

J. Stock. “Amperometric, Bipotentiometric, and Coulometric Titration” Anal. Chem, vol. 50, pp. 1-7, 1980. DOI: https://doi.org/10.1021/ ac50055a001

T. Michalowski, M. Toporek, M. Rymanowski. “pH-Static Titration: A Quasistatic Approach”. J. Chem. Educ, vol. 84, pp. 142-150, 2007. DOI: https://doi.org/10.1021/ed084p142

T. Lenoir, A. Manceau. “Number of Independent Parameters in the Potentiometric Titration of Humic Substances” Langmuir, vol. 26, 3998–4003, 2010. DOI: https://doi.org/10.1021/la9034084

A. Olin, B. Wallén. “On the accuracy of acid-base determinations from potentiometric titrations using only a few points from the titration curve” Talanta, vol. 24, pp. 303–308, 1977. DOI: https://doi.org/10.1016/0039- 9140(77)80006-4

W. Fish, F. Morel. “Propagation of error in fulvic acid titration data: a comparison of three analytical methods” Can. J. Chem, vol. 63, pp. 1185-1193, 1984. DOI: https://doi.org/10.1139/v85-202

R. Brown, M. Milton, P. Brewer. High Accuracy Titrimetry with Application to HCl. Londres: NPL Report COAM 5, 2001, p. 201. ISSN: 1475-6684

D. Harris. Quantitative Chemical Analysis: W. H. New York: Freeman and Company, 2006. P. 222

L. Schwartz. “Uncertainty of a titration equivalence point: A graphical method using spreadsheets to predict values and detect systematic errors” J. Chem. Educ., vol. 69, pp. 879, 1999. DOI: https://doi. org/10.1021/ed069p879

M. Pueyo, J. Obiols, E. Vilalta. “Expression of Uncertainty of an Acid-Base Titration” Anal. Commun, vol. 33, pp. 205-208, 1996. DOI: https://doi.org/10.1039/AC9963300205

International Organization of Standardization (ISO), ISO/IEC 17025: General requirements for the competence of testing and calibration laboratories, Genova: ISO, 2017

S. Ellison, A. Williams. “Quantifying Uncertainty in Analytical Measurement”, 2012. [En línea]. Disponible en: www.eurachem.org/ images/stories/Guides/pdf/QUAM2012_P1.pdf. [Último acceso: 31 07 2018]

Joint Committee for Guides in Metrology (JCGM). “Evaluation of measurement data – Guide to the expression of uncertainty in measurement” 2008. [En línea]. Disponible en: http://www.bipm.org/ en/publications/guides/gum.html. [Último acceso: 31 07 2018]

M.G. Cáñez-Carrasco, A.M. García-Alegría, M.A.G. Corella- Madueño. “Estimación de la incertidumbre expandida en titulaciones ácido-base” EPISTEMUS, vol. 18, No. 9, pp. 5-12N, 2015

B. Wampfler, M. Rösslein. “Uncertainty due to volumetric operations is often underestimated” Talanta, vol. 78, pp. 113–19, 2009. DOI: https://doi.org/10.1016/j.talanta.2008.10.047

International Organization for Standardization (ISO). ISO Guide 33:2015 - Reference Materials -- Good Practice in Using Reference Materials, Genova: ISO, 2015, p. 32

R Core Team. “R: A Language and Environment for Statistical Computing” 2018. [En línea]. Disponible en: https://cran.r-project.org/ doc/manuals/r-release/fullrefman.pdf. [Último acceso: 31 07 2018]

M. Cáñez-Carrasco, A. García-Alegría, A.T. Bernal-Mercado, R.A. Federico-Pérez, J. D. Wicochea-Rodríguez. “Conductimetría y titulaciones: ¿Cuándo, por qué y para qué?” Educ. quím., vol. 22, pp. 166-169, 2011

G. Gran. “Determination of the equivalence point in potentiometric titrations. Part II” Analyst, vol. 77, pp. 661-671, 1952. DOI: https://doi. org/10.1039/AN9527700661

M. Herrador, A. González. “Evaluation of measurement uncertainty in analytical assays by means of Monte-Carlo simulation” Talanta, vol. 64, pp. 415–422, 2004. DOI: https://doi.org/10.1016/j.talanta.2004.03.011

L. Schwartz, R. Gelb. “Statistical uncertainties of end points at intersecting straight lines” Anal. Chem., vol. 56, pp. 1487–1492, 1984. DOI: https://doi.org/10.1021/ac00272a063

Dimensions

PlumX

Visitas a la página del resumen del artículo

12416

Descargas

Los datos de descarga aún no están disponibles.

Cómo citar

[1]
“Técnicas de titulación ácido-base: consideraciones metrológicas”, Rev. Colomb. Quim., vol. 48, no. 1, pp. 26–34, Jan. 2019, doi: 10.15446/rev.colomb.quim.v48n1.72401.